Test Code TMSI Microsatellite Instability, Tumor
Necessary Information
A pathology report (final or preliminary), at minimum containing the following information, must accompany specimen for testing to be performed:
1. Patient name
2. Block number-must be on all blocks, slides, and paperwork (can be handwritten on the paperwork)
3. Tissue collection date
4. Source of the tissue
Specimen Required
This assay requires at least 40% tumor nuclei for endometrial specimens and at least 20% tumor nuclei for colorectal specimens.
-Preferred amount of tumor area with sufficient percent tumor nuclei: tissue 72 mm(2)
-Minimum amount of tumor area: 18 mm(2)
-These amounts are cumulative over up to 10 unstained slides and must have adequate percent tumor nuclei.
-Tissue fixation: formalin-fixed paraffin-embedded, non-decalcified
Preferred: Submit 2, if available, of the following specimens.
Acceptable: Submit at least one of the following specimens.
Specimen Type: Tissue block
Collection Instructions: Submit a formalin-fixed, paraffin-embedded tissue block with acceptable amount of tumor tissue.
Specimen Type: Tissue slide
Slides: 1 Hematoxylin and eosin-stained and 10 unstained
Collection Instructions:
Submit the followings slides:
1 Slide stained with hematoxylin and eosin
AND
10 Unstained, nonbaked slides with 5-micron thick sections of the tumor tissue.
Note: The total amount of required tumor nuclei can be obtained by scraping up to 10 slides from the same block.
Additional Information: Unused unstained slides will not be returned.
Forms
1. Molecular Genetics: Inherited Cancer Syndromes Patient Information (T519)
2. If not ordering electronically, complete, print, and send 1 of the following forms with the specimen:
-Gastroenterology and Hepatology Test Request (T728)
-Oncology Test Request (T729)
Useful For
Evaluation of tumor tissue to identify patients at high risk for having Lynch syndrome, also known as hereditary nonpolyposis colorectal cancer
Evaluation of tumor tissue for clinical decision-making purposes given the prognostic and therapeutic implications associated with microsatellite instability phenotypes
Disease States
- Lynch syndrome
Testing Algorithm
When this test is ordered, slide review will always be performed at an additional charge.
For more information see Lynch Syndrome Testing Algorithm
Special Instructions
Reporting Name
Tumor, Microsatellite InstabilitySpecimen Type
VariesSpecimen Stability Information
| Specimen Type | Temperature | Time |
|---|---|---|
| Varies | Ambient (preferred) | |
| Frozen | ||
| Refrigerated | ||
Reject Due To
All specimens will be evaluated at Mayo Clinic Laboratories for test suitability.Reference Values
An interpretive report will be provided.
Supportive Data
A total of 100 accuracy samples were run retrospectively during verification; 55/100 samples were colorectal, 41/100 were endometrial, and the remaining 4/100 were other tumor types. The overall concordance between the Idylla and Promega results was 98/100 (98%). Seventy-nine of 100 samples were microsatellite stable (MSS) by Promega and 77 (97%) had concordant MSS results by Idylla. Twenty-one of 100 samples were microsatellite instability-high (MSI-H) by Promega and 21 (100%) had concordant MSI-H results by Idylla.
In addition to the retrospective samples, 100 consecutive samples were prospectively analyzed, of which 58 were colorectal, 31 were endometrial, and 11 were from other tumor types. Seventy-six of 100 samples were MSS by Promega and all (100%) were MSS by the Idylla assay. Twenty-four of 100 samples were MSI-H by Promega. Twenty-three of 24 (96%) of the MSI-H samples were concordant by Idylla. One patient had an uncommon reason for testing. The discordant sample DNA was rerun on the Promega platform, but there was not sufficient tissue remaining to rerun this specimen on the Idylla assay. After reviewing the results from the 2 runs on Promega, a consensus decision amongst 5 pathologists was reached, and the sample was reclassified as equivocal by the Promega assay.
Precision and reproducibility were evaluated by running 3 MSI-H samples and 3 MSS samples in triplicate on the same instrument. Each of these samples had a 4th cartridge run on a separate instrument. There was 100% concordance between replicates from the 3 MSI-H samples and 3 MSS samples.
Day(s) Performed
Varies
Report Available
4 to 7 daysSpecimen Retention Time
Tissue blocks: Unused portions of blocks will be returned; Tissue slides: Unused slides are stored for at least 5 years; Extracted DNA: 3 monthsPerforming Laboratory
Mayo Clinic Laboratories in Rochester
CPT Code Information
81301
88381-Microdissection, manual
LOINC Code Information
| Test ID | Test Order Name | Order LOINC Value |
|---|---|---|
| TMSI | Tumor, Microsatellite Instability | 81711-4 |
| Result ID | Test Result Name | Result LOINC Value |
|---|---|---|
| 609365 | Result Summary | 50397-9 |
| 609366 | Result | 43368-0 |
| 609367 | Interpretation | 69047-9 |
| 609368 | Specimen | 31208-2 |
| 609369 | Source | 31208-2 |
| 609370 | Tissue ID | 80398-1 |
| 609371 | Released By | 18771-6 |
Additional Tests
| Test ID | Reporting Name | Available Separately | Always Performed |
|---|---|---|---|
| SLIRV | Slide Review in MG | No, (Bill only) | Yes |
Specimen Minimum Volume
See Specimen Required