Test Code CBBRP Coxiella burnetii (Q Fever), Molecular Detection, PCR, Blood

Performing Laboratory

Mayo Clinic Laboratories in Rochester

Reporting Name

Coxiella burnetii (Q fever) PCR, B

Specimen Type

Whole Blood EDTA


Ordering Guidance


 



Specimen Required


The high sensitivity of amplification by polymerase chain reaction requires the specimen to be processed in an environment in which contamination of the specimen by Coxiella burnetii DNA is unlikely.

 

Container/Tube:

Preferred: Lavender top (EDTA)

Acceptable: Royal blue top (EDTA), pink top (EDTA), or sterile vial containing EDTA-derived aliquot

Specimen Volume: 1 mL

Collection Instructions: Send whole blood specimen in original tube (preferred).


Reject Due To

All specimens will be evaluated at Mayo Clinic Laboratories for test suitability.

Specimen Stability Information

Specimen Type Temperature Time
Whole Blood EDTA Refrigerated (preferred) 7 days
  Frozen  7 days

Specimen Minimum Volume

0.5 mL

Day(s) Performed

Monday through Friday

Specimen Retention Time

1 week

Report Available

Same day/1 to 4 days

Reference Values

Not applicable

Useful For

Aiding in the diagnosis of Coxiella burnetii infection (eg, Q fever)

CPT Code Information

87798

LOINC Code Information

Test ID Test Order Name Order LOINC Value
CBBRP Coxiella burnetii (Q fever) PCR, B 90443-3

 

Result ID Test Result Name Result LOINC Value
35191 Specimen Source 31208-2
35192 Coxiella burnetii PCR 90443-3

Method Description

Bacterial nucleic acid is extracted from the specimen using the automated MagNA Pure instrument. The purified DNA is placed on the LightCycler instrument, which amplifies and monitors by fluorescence the development of target nucleic sequences after each PCR cycle. A specific target sequence from Coxiella burnetii is amplified and the resulting segment is detected using specific hybridization probes. Detection of the C burnetii target is performed through melting curve analysis using the LightCycler software.(Cockerill FR, Uhl FR: Applications and challenges of real-time PCR for the clinical microbiology laboratory. In: Reischl U, Wittwer C, Cockerill F, eds. Rapid Cycle Real-Time PCR, 2002:3-27; Kersh GJ, Bleeker-Rovers CP: Coxiella. In: Carroll K, Pfaller M, eds. Manual of Clinical Microbiology. 12th ed. ASM Press; 2019:1180-1188)

Supportive Data

This assay was clinically validated in a blinded manner using 52 archived, formalin-fixed, paraffin-embedded heart valve specimens from patients with endocarditis. A single sample within this set determined to contain polymerase chain reaction (PCR) inhibitors was omitted from the final analysis set. Compared with existing diagnostic data, PCR had a sensitivity of 100% (8/8) and specificity of 100% (43/43). All samples were assayed with a second PCR assay targeting the IS1111 element.(1) Complete concordance was noted between the 2 assays (P >0.999). The limit of detection of the assay is 2.16 targets/mcL for EDTA whole blood.