Test Code HMUOE Heavy Metal Occupational Exposure, with Reflex, Random, Urine
Specimen Required
Patient Preparation:
1. For the 48-hour period prior to start of collection, patient should not eat seafood.
2. High concentrations of gadolinium and iodine are known to potentially interfere with most inductively coupled plasma mass spectrometry-based metal tests. If either gadolinium- or iodine-containing contrast media has been administered, a specimen should not be collected for 96 hours.
Supplies: Urine Tubes, 10 mL (T068)
Collection Container/Tube: Clean, plastic urine container with no metal cap or glued insert
Submission Container/Tube: Plastic, 10-mL urine tube or clean, plastic aliquot container with no metal cap or glued insert
Specimen Volume: 6 mL
Collection Instructions:
1. Collect a random urine specimen.
2. See Metals Analysis Specimen Collection and Transport for complete instructions.
Useful For
Preferred screening test for detection of arsenic, cadmium, mercury, and lead due to occupational exposure using random urine specimens
Profile Information
Test ID | Reporting Name | Available Separately | Always Performed |
---|---|---|---|
ASOU | Arsenic Occupational Exposure | Yes, (order ASUOE) | Yes |
CDOU | Cadmium Occupational Exposure | Yes, (order CDUOE) | Yes |
HGOU | Mercury Occupational Exposure | Yes, (order HGUOE) | Yes |
PBOU | Lead Occupational Exposure | Yes, (order PBOUE) | Yes |
CRETR | Creatinine, Random, U | No | Yes |
Reflex Tests
Test ID | Reporting Name | Available Separately | Always Performed |
---|---|---|---|
SPAS | Arsenic Speciation, Random, U | Yes | No |
Testing Algorithm
If arsenic concentration is greater than or equal to 10 mcg/L, then speciation will be performed at an additional charge.
Special Instructions
Reporting Name
Heavy Metal Occ Exp w/Reflex, USpecimen Type
UrineSpecimen Minimum Volume
3 mL
Specimen Stability Information
Specimen Type | Temperature | Time | Special Container |
---|---|---|---|
Urine | Refrigerated (preferred) | 7 days | |
Frozen | 7 days |
Reject Due To
All specimens will be evaluated at Mayo Clinic Laboratories for test suitability.Reference Values
ARSENIC:
Biological Exposure Indices (BEI): <35 mcg/L at end of work week
CADMIUM:
BEI: <5.0 mcg/g creatinine
MERCURY:
BEI: <35 mcg/g creatinine
LEAD:
BEI: <150 mcg/g creatinine
CREATININE:
≥18 years: 16-326 mg/dL
Reference values have not been established for patients who are younger than 18 years.
Specimen Retention Time
14 daysPerforming Laboratory
Mayo Clinic Laboratories in RochesterCPT Code Information
82175
82300
83825
83655
82570
LOINC Code Information
Test ID | Test Order Name | Order LOINC Value |
---|---|---|
HMUOE | Heavy Metal Occ Exp w/Reflex, U | 29589-9 |
Result ID | Test Result Name | Result LOINC Value |
---|---|---|
CRETR | Creatinine, Random, U | 2161-8 |
608890 | Arsenic Occupational Exposure | 13463-5 |
608892 | Cadmium Occupational Exposure | 13471-8 |
608893 | Mercury Occupational Exposure | 13465-0 |
608894 | Lead Occupational Exposure | 13466-8 |
608891 | Total Arsenic Concentration | 5586-3 |
Day(s) Performed
Monday through Friday