Sign in →

Test Code FL Fluoride, Plasma

Additional Codes

FLORIDE

Performing Laboratory

Mayo Clinic Laboratories in Rochester

Reporting Name

Fluoride, P

Specimen Type

Plasma Heparin


Specimen Required


Supplies: Sarstedt Aliquot Tube, 5 mL (T914)

Collection Container/Tube: Green top (sodium heparin)

Submission Container/Tube: Plastic vial

Specimen Volume: 3 mL Plasma

Collection Instructions: Centrifuge and aliquot plasma into a plastic vial. Glass tubes are not acceptable.


Reject Due To

Gross hemolysis OK
Gross lipemia OK
Gross icterus OK

Specimen Stability Information

Specimen Type Temperature Time
Plasma Heparin Refrigerated (preferred) 14 days
  Ambient  14 days
  Frozen  14 days

Specimen Minimum Volume

Plasma: 1.2 mL

Day(s) Performed

Tuesday

Specimen Retention Time

14 days

Report Available

2 to 8 days

Reference Values

<4.1 mcmol/L

Useful For

Assessing accidental fluoride ingestion

 

Monitoring patients receiving sodium fluoride for bone disease or patients receiving voriconazole therapy

CPT Code Information

82735

LOINC Code Information

Test ID Test Order Name Order LOINC Value
FL Fluoride, P 14726-4

 

Result ID Test Result Name Result LOINC Value
8641 Fluoride, P 14726-4

Method Description

The fluoride electrode consists of a single-crystal lanthanum fluoride membrane and an internal reference bonded into an epoxy body. The crystal is an ionic conductor in which only fluoride ions are mobile. When the membrane is in contact with a fluoride solution, potential develops across the membrane. This potential depends on the level of free fluoride ions in solution and is measured against an external constant reference potential with a digital pH/mV meter.(Instruction manual: Fluoride Ion Selective Electrode User Guide. Thermo Scientific; 09/2016)